GE eBike III basic Series Service Manual page 73

Electric bike with and without nibp
Table of Contents

Advertisement

Select „Dynamic load calibration" and confirm with NIBP
Start the offset calibration process with NIBP
The line „DMS" shows:
stored DMS value / actual DMS value
By pressing NIBP (START) the actual value is averaged
for 5 sec and stored
Press NIBP (START) to start calibration step #1
The eBike III automatically sets the load to 50 Watt
Set the calibrator unit to 50 rpm
The line „RPM" shows the measured rotations/min
(issued by the calibrator unit)
[the value „P" is not used in the eBike III)
The line „Load" shows the target load in Watt-
automatically set by the eBike III (e.g. 50 Watt)
[the value „AW" is not used in the eBike III)
Wait approx. 20 seconds, until the load display is
stable
By using the arrow keys, input the actual measured
load value of the calibrator (e.g. 51,3 Watt)
Store the value with SET.
2018111-336 Rev. B
Service Menu
Contrast
ECG Type
Baudrate 8N2
Analog Calibration
Software update
Dynamic Load Calibr.
Saddle Calibration
S
election of parameterS in Service menu
Dynamic Load Cal.
Calibration offset
Gain:
DMS:
RPM:
Load:
d
ynamic load offSet calibration
Dynamic Load Cal.
Calibration offset
Gain:
DMS:
RPM:
Load:
d
ynamic load offSet calibration
Dynamic Load Cal.
Dyn. Calibration
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
At 50 /50 W
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
eBike III basic / comfort
1,00
3277 / 3219
0 B
0 P
0 SW
0 AW
START
-
Start
waiting for
1,00
3277 / 3222
0 B
0 P
0 SW
0 AW
WAITING...
-
averaging
1,00
3219 / 3219
0 B
0 P
0 SW
AW
START
1 -
Start
waiting for
51,3 W
1,00
3219 / 3219
50 B
0 P
50 SW
AW
SET
1 -
input load value
- 73 -