As soon as the self test displays on the ergometer LCD,
press the two arrow keys simultaneously to activate the
service menu again
Select service code "999" with the arrow keys and confirm
by pressing NIBP.
Select "ECG type" by using the arrow keys and confirm by
pressing NIBP
To prevent settings from being changed in error, the ECG
type menu is protected with a password.
Using the arrow keys, enter "003" and confirm the entry by
pressing NIBP.
Select the ECG type "Analog/Digital" and confirm the
selection by pressing NIBP.
Power off the ergometer to leave the service menu.
MaC 5500 / 5000 st
Please refer to the manual of the MAC 5500 / 5000 ST
system.
2018111-336 Rev. B
software settIngs
eBike III basic / comfort
Service Code
999
Service Menu
Contrast
ECG type
NIBP mmHg / kPa
NIBP Beep
Baudrate 8N2
Analog Calibration
Software update
ECG Type
003
ECG-Type
Digital
Analog / Digital
- 93 -