GE eBike III basic Series Service Manual page 69

Electric bike with and without nibp
Table of Contents

Advertisement

a
"I
: a
lIgnMent
nput
nalog
IN-Volt:
displays the voltage at the analog input
Set the ECG recorder output to a low load value
(for example 20 Watt).
The corresponding voltage is displayed (IN-Volt).
Wait until the voltage has settled.
Select "OFFSET" and confirm by pressing NIBP.
Use the arrow keys to set the load value to the ECG
recorder vaue (for example 20 Watt) and confirm by
pressing NIBP.
The setting is stored in the ergometer.
Set the ECG recorder output to a high load value
(for example300 Watt).
The corresponding voltage is displayed (IN-Volt).
Wait until the voltage has settled.
Select "GAIN" and confirm by pressing NIBP.
Use the arrow keys set the load value to the ECG recorder
vaue (for example 300 Watt) and confirm by pressing NIBP.
The setting is stored in the ergometer.
Setting range:
OFFSET: range from 20 to 1000 Watt in 10 Watt steps
GAIN:
range from 20 to 1000 Watt in 10 Watt steps
2018111-336 Rev. B
l
"
oad Control
eBike III basic / comfort
INPUT 1
Load
IN-Volt:
OFFSET
GAIN
EXIT
a
lignment of analog input
load
INPUT 1
Load
IN-Volt:
OFFSET
GAIN
EXIT
o
a
ffSet
lignment of analog input
INPUT 1
Load
IN-Volt:
OFFSET
GAIN
EXIT
a
lignment of analog input
load
INPUT 1
Load
IN-Volt:
OFFSET
GAIN
EXIT
g
ain alignment of analog input
0.10 V
20 Watt
300 Watt
"
0.10 V
20 Watt
300 Watt
"
load
3.05 V
20 Watt
300 Watt
"
3.05 V
20 Watt
300 Watt
"
load
- 69 -

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ebike iii comfort series

Table of Contents