3. Check cranks and pedals:
- pedals have no signs for damage / wear
- pedals securely fixed on cranks
- crank screws securely fixed (torque 40 Nm ± 2 Nm) on axis
Expected result:
eBike III basic
- Check bellow on saddle for damage / wear / cracks
- Check handlebar adjustment (angle)
- Check saddle height adjustment
- Check that the clamping levers (saddle [height] and handlebar [angle] adjustment) are tight.
Grease the thread of the saddle clamping lever.
(Use of an universal high-performance lubricant such as OKS 470 is recommended.)
eBike III comfort
- Check handlebar adjustment (angle)
- Check that the clamping lever is tight.
- Check electrical saddle adjustment: adjust minimum and maximum saddle height;
then return saddle to mid-level
- Check correct indication of saddle height display
- Check that the clamping lever for handlebar adjustment (angle) is tight
Expected result:
4. Using the power cord, connect the device to the mains power line and place the power switch in the ON position
Wait for self test to end:
- check LCD for error messages (refer to „Turning the System On" on page 60)
- check the software version in the service menu for correct version GF 1.x
(refer to „Software update" on page 78)
- check the "error log file" in service menu (refer to „Error log" on page 67)
Expected result: Device powers on without errors, FW is correct version, no errors logged
5. Check speed indication (rpm) is displayed on LCD and the additional speed indication on the patient display.
Check noise level of idling drive unit (for example, for grinding noises or noise from the bearing)
Expected result:
6. Units with NIBP module:
- Check cuff tubing and tubing connections, microphone connectors
- Check blood pressure measurement on test subject at rest (refer to the „eBike III operators manual")
- Check blood pressure readings are displayed on LCD
Expected result:
7. Set manual load (see „Manual Load" on page 66), change load (for example, 50 W, 100 W)
- Check load level and load indication
Expected result:
8. If ergometer is remote controlled from ECG recorder or PC-ECG system:
- Check connecting cable
- Check load control via ECG recorder or PC ECG system
- Check remote start if applicable
Expected result:
2018111-336 Rev. B
Cranks and pedals OK
Adjustments work and levers are tight, thread properly greased
Speed displayed on both displays and no noticeable noise
BP values are displayed and in proper range
Load is set and changes as expected
Ergometer can be controlled by external unit
eBike III basic / comfort
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail