GE eBike III basic Series Service Manual page 82

Electric bike with and without nibp
Table of Contents

Advertisement

B
III
e
Ike
software settIng
To activate the service menu, the ergometer has to be
powered off and on again.
As soon as the self test displays on the ergometer LCD,
press the two arrow keys simultaneously:
To prevent settings from being changed in error, the
service menu is "protected" with a password.
Select "999" with the arrow keys and confirm by pressing
NIBP.
Select "ECG type" by using the arrow keys and confirm by
pressing NIBP
To prevent settings from being changed in error, the ECG
type menu is protected with a password.
Using the arrow keys, enter "003" and confirm the entry by
pressing NIBP.
Select the ECG type "Digital" and confirm the selection by
pressing NIBP.
Power off the ergometer to leave the service menu.
Case
software settIngs
Please refer to the manual of the CASE system.
- 82 -
Contrast
ECG type
NIBP mmHg / kPa
NIBP Beep
Baudrate 8N2
Analog Calibration
Software update
ECG Type
ECG-Type
Digital
Analog / Digital
eBike III basic / comfort
Service Code
999
Service Menu
003