GE AxiChrom 50 Operating Instructions Manual page 125

Columns
Table of Contents

Advertisement

3 Gently pull on the distributor to withdraw a length of the tube [405-1] from inside
4 Unscrew the adapter tube [405-1] from the top distributor [402] by hand.
5 Pull out the adapter tube [405-1] from the top end of the adapter rod [802]. Unscrew
6 Remove the scraper seal [605] and O-ring [604] by gently moving them up the dis-
7 Check the scraper seal for scratching and wear, and the O-ring for twisting and wear.
8 Rinse the distributor with water and then inspect for signs of damage. If necessary
9 Check the adapter plate O-ring [127] for visible signs of damage. If damaged replace
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
the adapter rod [802].
Figure 6.12: Withdrawing a length of the tubing from inside the adapter rod.
Figure 6.13: Unscrewing the adapter tubing from the distributor.
the tube connection [109] from the adapter tubing [405-1]. Check the tubing and O-
rings [602a, b] for visible signs of damage. Replace if necessary.
tributor a bit at a time around the circumference.
Note:
The O-ring is secured within the scraper seal so they will most likely be
removed together.
If damaged or if there is any doubt about their integrity, discard them.
replace the distributor.
it with a new one.
6 Maintenance and service
6.2 Performing maintenance and service
125