5 Click on Next to display the Slurry and Packing Test Parameters wizard page. Make
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
5.1 Intelligent Packing of the column with ÄKTAexplorer and ÄKTApilot
the appropriate selections.
Figure 5.4: Intelligent Packing Wizard (4 of 5), ÄKTApilot
Slurry concentration
Equilibration volume
Use default liquid velocity
during equilibration
Use default liquid velocity
during packing test
Packing test upflow and
Packing test downflow
Enter the slurry concentration (see Preparing the
slurry, on page 63).
Enter an Equilibration volume (measured in CV)
within the designated range. Default values are
given according to the recommended packing and
test liquid for the selected medium.
Leave checked to use the suggested default liquid
velocity of 30 cm/h.
Leave checked to use the suggested default liquid
velocity of 30 cm/h.
Select whether to perform packing tests upflow,
downflow or both. Both are selected by default.
5 Operation
69