4.8 Preparing the column for a run
Moving a column with an AxiChrom pivot stand
The pivot stand facilitates easy repositioning or moving of the column.
A
Figure 4.6: Moving the column using the pivot stand. Hold the column by the stand arm
and/or the top bayonet (A), and NOT on the adapter rod (B).
1 Tilt the column backwards onto the wheels.
2 Hold the column by the stand arm and pull the column backwards. Pulling the column
If transporting the column in the pivot stand without the top plate and adapter [A] fitted,
place a centering tool into the mouth of the column to secure the tube in place. For
correct positioning of the centering tool, see Figure 6.26. After moving the column to its
resting position, level the column (see below).
54
is more stable than pushing it.
Note:
It is recommended to use the mechanical locking equipment when moving
a packed column.
CAUTION
Make sure not to touch the locking equipment for the arm of
the pivot stand when moving the column, to avoid injury and
rotation of the column.
B