Table 6-1.
HP-IS
Interface
Electronics System
Troubleshooting
(Cont'd)
DATA
ERRORS
(Cont'd)
SYMPTOM
CAUSE
ACTION
Error in displayed
1. Compensation
1. Check output of
reading
i s
less
circuitry
compensation cir-
than one part in
malfunction
cuitry.
100,000 and is evident
on all measurement
axes.
Laser system reads
Misalignment of the
2. If Comp. number
a shorter distance
optics resulting in
correct, check
than the known
cosine error.
for cosine
movement of the
error in optical
optics.
alignment (Refer
to Section II, Laser
and Optics).
3. If Comp. number
incorrect, re-
place 10755. If
using 5510, check
BCD inputs to
10755 using logic
probe and verify
correct Comp.
number is being
input.
Error in displayed
1. 10760A Mal-
1.
Troubleshoot
reading that is
function.
or replace 10760A
less than one
part in 100,000
and is evident on
2. Cosine error
2. Align optics
a single measure-
in optical
(Section 11)
ment axis.
alignment
3. 107806 Saturation
3. Adjust 10780A
(Refer to
10780A Manual)
ERROR
DETECTION
4. If an error indication does not occur when the laser beam is interrupted in a measurement axis:
a)
Check
LED
indicator
on
associated
10780A
receiver.
If
it
extinguishes
when
beam
is
interrupted,
proceed
to
step
b.
If
it
remains
on
when
beam
is
interrupted,
1.
Monitor
10780A
test
point
on
back
of
receiver.
If
voltage
is
greater
than
1.1V
dc,
refer
to
10780A
manual
to
reduce
receiver
gain.
2.
If
voltage
is
in
limits,
check
for
incorrect
rotation
of optics
or
receiver
(refer
to
Section
11).
3. If problem continues, troubleshoot or replace 10780A receiver.
b) If error does not register on any measurement axis, check, replace or troubleshoot system
interface card (10745).
c) If error does not register on one measurement axis only, replace or troubleshoot the associated
10760A card.