HP 5501A Operating And Service Manual page 232

Laser transducer system
Hide thumbs Also See for 5501A:
Table of Contents

Advertisement

Table
6-2.
Binary
Controlled
Counter
Interface
Electronics
Systems
Troubleshooting
(Cont'd)
CHECKOUT
PROCEDURE
I
REQUIRED
TEST
EQUIPMENT
INSTRUMENT
TYPE
REQUIRED
CHARACTERISTICS
Digital
Voltmeter
Range:
-15V
to
+15V
+
.01V dc
Logic
Probe
TTL
Level
Oscilloscope
Ability
to
display
signals
between
dc
and
20
MHz
Individual
Circuit
Card
Operating
and
Service
Manuals
MULTI-AXIS
PROBLEM
1 . I f same problem exists on all axes, check the following possible error sources
a) Check temperature in the coupler environment.
COUPLER
TEMPERATURE
SYMPTOMS
ACTION
O0
to
40°C
None
-
Electronics
None
should
operate
a s
described
Out
of
range
Counters
may
jump
or
Ventilate
count
erratically
enclosure
b) Check voltage on each coupler card.
LOCATION
VOLTAGE
+5V Front-edge test point
At least +5V
2
.01V on
each card.
C)
Check user software (Refer to Section IV for example Comparator Program.)
d)
Check instruction sequence timing (Refer to 10746 manual.)
e)
10746A
Binary
lnterface
Card
I
I
SYMPTOM
Data LED does not
light when con-
troller sends com-
mand to transfer
data
In normal operation,
Data LED i s dimly
lit (If sampling
continuously.)
POSSIBLE PROBLEM
I
ACTION
1. User software
2. 1/O Cabling
Refer to Section IV
for instruction
sequence.
Verify proper
connections and no
shorting.
3. Controller
malfunction
Run controller
diagnostic.
4. 10746A Mal-
Replace or Trouble-
function.
shoot 10746A.