The 9800 series calculator serves a s a system controller (Figure 3-2). The reference and mea-
surement signals from the laser head and receivers are compared continuously by the 10760A
counters. These signals are approximately 2 MHz and are driven by line drivers to ensure
noise immunity. Upon command from the calculator, the accumulated counts for each mea-
surement axis can be sampled independently or simultaneously. Pulses are generated and
accumulated in the counters only when movement has occurred. Refer to Table 3-7 for typical
system specifications.
The
10755A
compensation
interface
is
used
with
either
the
10756A
manual
compensator
or
the
5510A
Opt
010
automatic
compensator.
In
typical
installations
the
choice
of
manual
or
automatic
compensation
is
determined
by
the
stability
of
the
ambient
environment.
The 10783A numeric displays can be used to display the relative movement on each axis. The
displays are interfaced to the 9800 series calculator via the HP-IB and receive and display informa-
tion only on calculator command. (Refer to the individual unit theory of operation for addi-
tional information.)
Table
3
Typical
HP-IS
lnterface
Electronics
System Specifications
Number
of
Measu
figuration
restrictions.
Measurement
Resdution/Mwimum
Allowable
Velocity:
The
measurement
reso-
lution and
velocity
depend
on
the
type
of
interferometer
used.
107024
Linear
lnterferometer
10705A
Single
Beam
lnterferometer
Resolutjon
h/4
A/24
Maximum
Velocity
q 1 2 )
30.q1.2)
51(2)
mm/sec
(in/sec)
10706A
Plane
Mirror
lnterferometer
3.5
Computer lnterface Electronics
The Computer lnterface Electronics for the Laser Transducer System is designed to interface
a wide variety of digital processors. The individual components are selected and assembled
by the user for a specific measurement application.
The following discussion subdivides the Computer lnterface Electronics into two typicai
types of systems; a 10760A Counter-based system and a 10762A Comparator-based system
(closed-loop).