Table 6-4. English/Metric Pulse Output Electronics System Troubleshooting (Cont'd)
3.
If
problem continues, troubleshoot or replace 10780A receiver.
b) If the error does not register on one measurement axis only, replace or troubleshoot the
associated 10760A card.
C)
If the controller does not recognize an error occurring on any axis, verify the error output lines
from the coupler are being monitored by the controller. The error line
i s
an output from
the 10763A.
FREQUENCY C O U N T E R
Table 6-4a : 5300/5301 Control Settings for Coupler Operation Verification
SET FRONT PANEL CONTROLS AS FOLLOWS:
GATE
OPEN
WAVEFORM
n
COUNTER INPUT LINE
CONNECT TO PIN
4
O N
FRONT-EDGE TEST FIXTURE
GROUND FROM INPUT BNC
CONNECT T O PIN 18 O N
FRONT-EDGE TEST FIXTURE
SAMPLE RATE
HOLD
7
Table
6-4b:
5328 Universal Counter Settings
SET FRONT PANEL CONTROLS AS FOLLOWS:
FUNCTION
START A
FREQ. RESOLUTION/N
1 MHz, 1
SAMPLE RATE
HOLD
(Use channel A input)
COUNTER INPUT LINE
CONNECT TO PIN 4 O N
FRONT-EDGE TEST FIXTURE
GROUND FROM INPUT BNC
CONNECT TO PIN 18 O N
FRONT-EDGE TEST FIXTURE
SLOPE
+
AC/DC
DC
AlTEN
1
LEVEL A
ADJUST TOWARD
+
SO THAT
THE COUNTER DISPLAY
STABILIZES ( N O COUNTING
SHOULD TAKE PLACE IF
LEVEL IS INCREASED.)
SEP/COM A
SEP