HP 5501A Operating And Service Manual page 227

Laser transducer system
Hide thumbs Also See for 5501A:
Table of Contents

Advertisement

Table
6-2.
Binary
Controlled
Counter
Interface Electronics
Systems Troubleshooting
(Cont'd)
e) 10746 Binary lnterface card.
SYMPTOM
POSSIBLE PROBLEM
ACTION
Data LED does
1. User software
Refer to Section
not light when
IV for instruction
controller sends
sequence.
command to trans-
fer data
2. 1/O cabling
Verify proper
connections and no
shorting.
3. Controller mal-
Run controller
function
diagnostic
4. 10746 mal-
Replace or trouble-
function
shoot 10746.
f)
Check
I/O cabling from controller
to
coupler.
g)
If
all
the
reference
or
measurement
LED
indicators
on
10760A
counter
cards
do
not
light
when
power
is applied
:
ACTION
LOCATION
RESULT
SUBSEQUENT ACTION
Verify presence
10760A front-
Signal
If LED is extin-
of reference
edge
present
guished, trouble-
and measure-
shoot 10760A
ment signals
with oscillo-
Measurement
Signal
Troubleshoot
scope. Signals
Signal: Pins
absent
receiver or
are square
18,20
cabling.
waves
Reference :
Signal
If LED extinguished,
pins 22,24
present
troubleshoot 10760A
Signal
Troubleshoot laser
absent
head or cabling.
PROBLEM
EXISTS
BUT
NOT
O N
ALL AXES
2.
If
the
system
is
multi-axis
but
the
problem
is
not
present
on
all
axes,
substitute
known
good
components
to
isolate
the
problem
system
component.
The
following
may
also
apply:
I
a) Check user software (refer to Section IV for example program).
b) Check voltage on each coupler card.
LOCATION
VOLTAGE
+5V Front-edge test point
At least +5V
+
.0lV
on each card.