Hitachi PRIMAIDE 1440 Instruction Manual page 116

Table of Contents

Advertisement

Primaide 1440 FL
Detector
Distilled water
Flow cell
Primaide 1110
Pump
Fig. 4-1 Tubing Diagram
(2)
Measurement of Raman Spectrum of Water
Set the flow cell unit in place, measure the Raman
spectrum of water under the conditions given below, and
record the results (see sections 3.4.2, 3.4.3, 3.4.5,3.4.6 and
3.5.2). Measure in the following sequence.
• Excitation (Ex) wavelength : 350 nm
• Time constant (response) : 2 seconds
• Photomultiplier voltage
• Emission spectrum
bandwidth
• Emission spectrum
measuring wavelength
• Spectrum No.
(a) Turn ON the power switch and the 1
appears.
(b) Set an excitation wavelength.
(i)
Press the
screen appears. Set an excitation (Ex) wavelength.
4 - 3
D-2500 Chromato-data processor
Waste
solution
: MID
: STANDARD (15 nm) or
WIDE (30 nm)
: 350 to 450 nm
: Any of 1 to 4
st
WL
key and the wavelength setting
monitor screen