2. Installation Conditions

Confirm that the following requirements are satisfied before
installing the instrument.
(1)
Power Supply
Power supply
voltage
Frequency
Power capacity : 250 VA or more for a single instrument,
Power outlet
APPENDIX 1 - 2
: 100 to 240 V AC
Fluctuation should be within 10% of the
rated voltage.
: 50 or 60 Hz
Fluctuation should be within 0.5 Hz of
the rated frequency.
and 850 VA or more for the total system.
The total system shown comprises
Primaide organizer, Primaide 1110 pump,
Primaide 1210 autosampler, Primaide
1310 column oven, and Primaide
1410/1430/1440 detector.
Check and prepare a power supply
meeting the power supply requirements
for your actual system configuration and
OA equipment to be used in combination.
: A power outlet shall be available near
the installation position of the system
including this instrument.
This instrument requires a single
grounded power outlet (3-pole type).
At least one additional outlet is needed
for the entire system.