5.3 Fuse Replacement

Electric Shock due to Contact with Inside of
Instrument
When replacing the fuse with a new one, be careful not to
receive an electric shock. To prevent this, be sure turn off
the power switch and unplug the power cord before
proceeding to fuse replacement.
Item to prepare:
• Item to prepare: 2 new fuses (P/N J821399 T5AH)
(1)
Items to Check: Prior to fuse replacement, make sure the
power switch of the instrument is off and the power cord is
unplugged.
(2)
Replacement of Fuse
(a) Hold the tabs on both sides of the fuse holder and
detach the holder.
Fig. 5-14
(b) Replace two fuses with new ones, then reattach the
fuse holder in place.