Create a time program as indicated below for measuring the
Raman spectrum of water.
TIME
Ex
0.0
350
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.4
1.5
1.6
1.8
2.0
NOTE:
4 - 13
Em
BASE (A/Z=1, HOLD=2)
450
1
360
370
375
380
385
390
395
400
405
410
415
430
450
The above time program shows an
example where the minimum FL value
was obtained at 380 nm and the
maximum at 400 nm. For the top Em of
the program, set a wavelength that
includes the wavelengths where the
maximum and minimum FL values were
obtained. (For instance, if the maximum
FL value was obtained at a wavelength
of 397 nm, then the wavelength at 1.1
minutes should be set at 397 nm.)