<Xe lamp energy check screen> (energy check is in progress)
CHECKING LAMP ENERGY
(4)
The result of Xe lamp energy check will be displayed
automatically.
<Xe lamp energy check result>
Xe ENERGY
1200
Press the ENT or ESC key.
* If 5 minutes or more elapses without pressing ENT or
ESC key, the system will return automatically to the 1st
monitor screen.
(5)
When the Xe lamp energy check is finished, the system
returns to the 1st monitor screen.
NOTE:
When the temperature of the Xe lamp is high, the
sealed gas pressure will fluctuate and the lamp may be
difficult to ignite. Before restarting the instrument or
re-lighting the lamp, wait at least five minutes for the
lamp to cool down after turning it off.