Hitachi PRIMAIDE 1440 Instruction Manual page 187

Table of Contents

Advertisement

Wash fluid
Primaide 1000
Organizer
Primaide 1110
Pump
Degassing
device
(2) Isocratic System(stacked in two blocks)
Mobile phase
Drain
Appended Fig. 1-8 Isocratic System
Connect the inlet tube of the mobile-phase container to the
IN port of the degassing device. Connect the OUT port of
the degassing device to the inlet of the pump. Connect the
drain valve of the pump to port 1 of the injection valve of the
autosampler. Connect port 2 of the injection valve to the
column inlet. Connect the column outlet to the inlet tube of
the detector cell. For detergent, provide connection to the
three way solenoid valve of the syringe valve of the
autosampler through the degassing device.
APPENDIX 1 - 15
Column
1
2
6 3
1
Drain
Primaide 1440
Fluorescence detector
Detector cell
Primaide 1310
Column oven
Primaide 1210
Autosampler