1540
1541
This error is logged when the message Data
transmission
abnormal termination.
appears.
The error is received from the remote system, and
the Status field on the printout shows the error
passed by that system.
This error is logged when the message Data
transmission
abnormal termination.
appears
and indicates a transmission problem at the local
cardiograph. Try each of these possible remedies:
1. Check cabling
2. Retry the operation
3. Use a different phone line
4. Check with the system operator to find out if
abnormalities exist at the other end
9000-9499
Errors in this range indicate a communication
problem between the cardiograph and the patient
module. The problem can be caused by hardware
or software, and can occur in the cabling, the CPU
assembly, or the patient module. To locate and
correct the problem, use these actions:
1. Check all cables
2.
Cycle the power
3. Run Extended Self-test
4. Reload the software
5. Refer to the troubleshooting table for the patient
module and cable
Using Extended
Self-test
in
Troubleshooting
The cardiograph's Extended Self-test is a simple-to-use, looping test
that gives you pass/fail status for each of these major subassemblies:
n
.Patient module and cable
n
CPU assembly
H Printer
H Disk drive (M1700A and M1701A only)
n
Preview display (M1700A only)
H Keyboard display
n
Modem and cable (M1700A only)
n
Fax modem (Ml 75GA only)
Instructions for entering and using Extended Self-test are found in
Chapter 4, "Performance Verification and Maintenance."