Care And Cleaning; Cleaning The Cardiograph And Patient Module; Cleaning The Keyboard - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Care and Cleaning
The cardiograph and its accessories are chemically resistant to
,-
common cleaning solutions and non-caustic detergents. The following
are some approved cleaning solutions for all of the outside surfaces of
the cardiograph, the patient module, and the cart.
n
Mild soap and water
n
Chlorine Bleach (3Oml/l water)
n
90% Isopropyl Alcohol
If the keyboard requires cleaning, you can remove it from the
cardiograph and immerse it in lukewarm water.
Cleaning the Cardiograph
and Patient Module
1. Unplug the power cord and ensure that the cardiograph is set to
Standby at the front panel.
2. Wipe the external surfaces of the cardiograph, the patient module,
and reusable electrodes with a soft cloth dampened in any of the
approved cleaning solutions listed above. Avoid the leadwire
inputs and cable connections.
Caution
Do not use any strong solvents or abrasive cleaning materials.
Do
not spill any liquids on the surface of the cardiograph.
Do not immerse the patient module.
Do not use any of the following to clean the cardiograph:
n
Ammonia-based cleaners
n
Acetone
n
Iocline-based cleaners
n
Phenol-based cleaners
n
Ethylene oxide sterilization (ETO)
n
Autoclave
Cleaning the Keyboard
1. Carefully insert a small flat-blacled screwdriver into the notches at
the top or bottom edge of the keyboard and pry the keyboard up.
2. Rinse off the keyboard and thoroughly dry it.
3. Snap it back into the cardiograph.
Caution
Do not wash the keyboard in hot water. Do not attempt to clean the
keyboard in a dishwasher.
4-10
Performance
Verification
and Maintenance