Installation And Configuration; Setting Up The Patient Module - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Installation
and
Configuration
VCG recording requires that you use the Frank lead set in addition
to the standard 12-lead set to obtain the VCG from the patient. The
following information describes how to set up the patient module and
attach leads to the patient for VCG recording.
Setting Up the Patient
To set up the patient module, perform the following steps.
Module
1. Turn off the cardiograph.
2. Replace the four shorting plugs on the patient module with Frank
lead wires I, E, M, H. Optionally, you may replace V4 (C4) and
V6 (C6) chest lead wires with Frank lead wires C and A.
3. Turn on the cardiograph.
The following tables describe how to set up the patient module for
the lead set you are using.
Table
D-l.
AHA Lead Set and Patient Module Configurations
Lead
I
E
c,
A
h/I
II
Lead Color Code
Label Color
Orange/Red
Orsnge/Yellow
Orange/Green
Orange/Brown
Orange/Black
Orange/Violet
Red
Yellow
Blue
Violet
Black
Violet
Table D-2. IEC Lead Set and Patient Module
Lead
Lead Color Code
I
Light Blue/Red
E
Light Blue/Yellow
C
Light Blur/Green
A
Light Blue/Brown
M
Light Blue/Black
II
Light, Blue/Violet
Slot Label
V4R
V3R
v4
VG
v7
H
Label Color
1 Slot Label
1
Grey
Pink
Brown
Violet
Orange
Violet,
~ C4R
C3R
C4
C6
c7
H
D-2
Vectorcardiography