Table D-5. Troubleshooting
(continued)
Symptom
Cause
What to do . . .
No VCG report is printed and the
One or more X, Y, Z leads are
Be sure that all Frank lead wires
message Defective
VCG data
missing or the ECG line is flat.
are correctly plugged into the
appears briefly on the cardiograph's
patient module.
display. The Electrode off message
appears on the Auto ECG report.
No VCG report is printed after the
Incorrect 12-lead Auto ECG report
Selecf.either a 3 x 4, 1R format
la-lead Auto ECG report.
format selected
with an X, Y, or Z lead as rhythm
strip or select a 3 x 4, 3R format
with X, Y, and Z leads as rhythm
strips.
VCG report format not selected to
In global configuration,
change the
enable VCG printing.
answer to the VCG? question to one
of the four possible VCG report
formats: VCGl, VCG2, VCG3, or
VCG4.
No VCG report is printed from the A VCG format was not selected to
In global configuration,
change the
stored VCG file.
enable VCG printing
answer to the VCG? question to one
of the four possible VCG report
formats: VCGI, VCG2, VCG3, or
VCG4.
No X, Y, or Z lead ECG data has
VCG report printing is impossible.
been recorded.
X, Y, Z data was recorded in
Analyze the ECG on a cardiograph
special mode (500 samples/second)
that is configured for VCG
on a cardiograph not configured for recording.
VCG recording.