ECG Simulation
Comments:
Record an ECG using an ECG simulator. Verify the following:
0
Trace activity for all 12 leads
0
No gross distortion of complexes or calibration pulses
0
Calibration pulses are of proper duration (200 ms) and amplitude
(1 mv>
The trace will vary depending on simulator settings used and
simulator type. Calibration pulse measurements will vary depending
on the cardiograph gain and speed settings.
Performance
Verification
and Maintenance
4-15