HP PageWriter XL M1705B Service Manual page 165

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Warning
0
When the cardiograph
is connected
to AC power, there are dangerous
voltages
on the line module terminals,
even when the AC switch is off.
When the AC switch is on, there are also dangerous
voltages
on the
terminals
of the mains transformer.
If you must have the AC power
connected
while the cardiograph
Is open, do not touch any of the
exposed
connectors.
The rectifier
assembly
capacitors
store hazardous
amounts of energy
during AC and battery operation.
Be careful not to touch any of the
exposed
capacitor
connections.
5, Remove the user interface assembIy as described earlier in this
chapter.
6. Remove the three screws from the rear panel as shown in
Figure 7-13.
Figure 7-13. Opening and Removing
the Top Cover Assembly.
7. Slide the top cover forward about 1 inch, then carefully lift it
straight up until there is enough clearance to tilt the left side up.
To remove the top cover assembly, continue as follows:
8. If the cardiograph is equipped with a preview display, disconnect
the cable that connects the preview screen to the preview logic
board and a ground wire. Disconnect the.four cables and the
ground wire that connect the top cover assembly to the CPU
assembly.
9. Lift the top cover assembly out of the way and set it aside.
7-18
Removal and Replacement

Hide quick links: