Software Installation; Vcg Configuration (A.03.Xx Software Only) - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Software
Installation
Note
I
vi?
Version A.03.1X software must be installed in the cardiograph prior
to installing VCG software.
To install the M1790A VCG software, perform the following steps.
1. Press [On-Standby) to place the cardiograph in Standby.
2. Insert the application disk, labeled "VCG Disk," in the disk drive
in the front of the cardiograph.
3. Turn on the cardiograph. The message Load VCC Application?
will appear.
4. Press Yes. When installation is complete, the following display
appears:
Auto
3x4
Report
Format
01-26-92
01:23PM
VCG Configuration
To configure the PageWriter XLi cardiograph for VCG recording,
(A.OS.XX Software
Only)
perform the following steps.
1. Press (Menu) twice until the following display appears.
Transmit
Store
,
Config
CheckDisk
Files
2. Press Config.
The slain configuration menu will appear.
Global
Configuration
ID
Transmit
Files
Exit
3. Press Global
for global configuration selections.
4. Select the VCG choices as shown in Table D-3.
5. Press Exit
twice when you have finished configuring your
cardiograph.
Table D-3 lists global configuration parameters and choices you need
for VCG recording.
Vectorcardiography
M1790A
D-3