Troubleshooting - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Manual Report Formats
When the cardiograph is configured for VCG recording, some manual
report formats with Frank leads are automatically
added to the list of
formats from which you can choose. One of the manual S-lead ECG
report formats contains the X, Y, and Z leads. Another format is
the custom 3-lead group, which can be set in the global cardiograph
configuration to include any combination of three out of
15
leads.
If you configure a custom S-lead group, you must choose the X, Y,
and Z leads in that sequence to be able to print a VCG report from
that format. The VCG report cannot be printed unless the leads are
presented in the correct sequence.
Caution
al
Troubleshooting
This chapter contains information
about troubleshooting
the VCG
_.
recording.
Table D-5. Troubleshooting
Symptom
Cause
What to do . . .
Extra lead labels VXl, VX2, VX3,
Research leads instead of Frank
In global configuration,
under
VX4 appear in the cardiograph's
leads are configured
"Special applications",
change the
display
answer to the Research leads?
question to "No" to enable Frank
leads. Turn the cardiograph off and
on again.
Extra lead labels V3R, V4R, and
The Frank lead wires are not
Be sure that all Frank lead wires
V7 instead of X, Y, and Z appear
correctly plugged into the patient
are correctly plugged in. Be sure
on the cardiograph's display
module. The H slot contains a
that all of the other leads are
shorting plug.
plugged into the patient module
and that no shorting plugs are
installed.
Turn the cardiograph off and on
again.
D-10
Vectorcardiography
Ml790A