Saecg (M1754A Only); Editing And Printing A Stored Saecg - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Note
.I
!?
Editing and Printing a Stored SAECG
To edit or print a stored SAECG report:
1. Press m
until the following display appears:
Print
Edit
Config
2. Press Edit . A display similar to the following appears:
Edit
ID#:
12345678
Select
Next
Previous
Exit
3. Press Next or Previous
until the patient ID number you want
appears.
4. Press Select . The onset/offset adjust menu sl~own earlier
appears.
If you know the patient ID number of the ECG you want to edit or
print, you can select the ECG as follows:
H Press More, then Find.
H Type the patient ID number. Press Enter.
The most recent ECG
with that patient ID number appears.
n
If you want a different ECG for that patient, you must scroll
through the list by pressing Previous .
n
When the ID number you want appears, press Select .
You can now adjust the onset and offset and print the report as
described earlier.
Deleting
a Stored
You can delete SAECGs from the main XLi menu by selecting the
SAECG
SAECGs to be deleted and pressing Delete.
B-10
M1754A Signal Averaged
ECG (SAECG) System