2.2.4 Manifold input lines

The following instructions provide the procedural steps to install the manifold. The
meter run manifold high (+) and low (-) pressures terminate in the transducer's
(EXIMV) + and – differential port cells. The differential port holes are located on
the bottom of the EXIMV (see Figure 2–4).
2.2.4.1
-
-
2.2.4.2
1) Install the isolation valves on the meter run.
2) Install the tubing between the meter run and the XFC. It should be noted
3) Leak check all connections.
Page 2–6
Materials not supplied
Stainless steel tubing
Tubing fittings
A backup wrench should always be used when working with
stainless steel tubing and valves. This prevents the fitting from
turning and/or putting tension on stainless steel tubing.
Figure 2–4 G4 EX EXIMV ports
Step-by-step instructions
that the fittings are not supplied with the unit.
Leaks in the tubing or manifold will introduce errors when
calibrating the transducer (EXIMV).