Toshiba ST-B10 Series Maintenance Manual page 92

Hide thumbs Also See for ST-B10 Series:
Table of Contents

Advertisement

5. DIAGNOSTIC TEST OPERATION
5.4.5.1 Drawer Test
In this test, the drawer operation is checked.
(1) Operation procedure
1. Select the Drawer Test(1) button on the I/O Communication Test Menu screen.
2. Click the Drawer1 Open(1) or the Drawer2 Open(2) button.
3. Make sure that Drawer Status on the screen is matched with the current condition of the drawer.
When the drawer is open, the Drawer Status should indicate "open".
When it is closed, the Drawer Status should indicate "close".
4. When the drawer performs correctly, click the OK button to terminate the test. If not, click the
NG button.
(2) Error code
Result
Error Code
0
Except 0
(3) Corrective action
Error
Drawer open error
<< Preliminary version >>
Error Code Ex
0
0
• The drawer may be damaged. Replace the drawer.
0
The drawer performs correctly.
0
Drawer open error
Cause & Measure
5- 30
5.4 Serviceman Diagnostic Test Program
Remarks
EO18-12017