Emulation Module Built-in Commands
The emulation module has some built-in commands (sometimes called
the "terminal interface") which you can use for troubleshooting.
You can enter the built-in commands using:
• A telnet (LAN) connection
• The Command Line window in the Emulation Control Interface
• A "debugger command" window in your debugger
To telnet to the emulation module
You can establish a telnet connection to the emulation module if:
• A host computer and the logic analysis system are both connected to a
local-area network (LAN), and
• The host computer has the telnet program (often part of the operating
system or an internet software package).
To establish a telnet connection:
1 Find out the port number of the emulation module.
The default port number of the first emulation module in an
HP16600A/700A series logic analysis system is 6472. The default port
of a second module in an HP16600A-series system is 6476. The default
port numbers of a third and fourth module in an expansion frame are
6480 and 6484. These port numbers can be changed, but that is rarely
necessary.
2 Find out the LAN address or LAN name of the logic analysis
system.
3 Start the telnet program.
If the LAN name of the logic analysis system is "test2" and you have
Solutions for CPU32
Chapter 13: Troubleshooting the Emulation Module
Emulation Module Built-in Commands
269