Chapter 3: Connecting and Configuring the Analysis Probe
Configuring the Analysis Probe
To set the ID switches
The HP E2480A provides an identification (ID) which may be used by
other system components. The ID consists of primary and secondary
values. The primary value is fixed (identifies CPU32 family) by
hardware. The secondary ID is set by the 8-bit switch on the analysis
probe, which must be configured to match the microcontroller being
used. Positions 1 - 7 of the switch generate a binary value which must
correspond to the last two digits of the microcontroller (binary 32 for
MC68332). Position 8 is reserved and should be set to the "1" position.
The figure below shows the switch settings for the MC68332.
Switch Settings for MC68332 Target System
78