Siemens SIPROCESS GA700 Operating Manual page 147

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Table 6- 23
Function assignment at function (pre-selection): "Other functions"
Function
Trigger
Switch pump
Measuring protection
Ack. curr. mess.
Software reset
"Level-triggered": Activation by rising signal edge, deactivation by falling signal edge. Active at signal level "High".
1)
"Edge triggered": Activation by rising signal edge.
2)
Requirement: optional gas management module is installed.
3)
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Remark / effect
Switch sample gas pump
The pump is switched on at a positive edge and switched off at a negative
edge.
Positive/negative
edge
1)
Activate measuring protection.
As long as the measuring protection is active, no other user actions on the
device are possible. This especially includes changes to the configuration as
Positive/negative
well as initialization of calibration via a digital input or via a bus interface.
edge
1)
A cyclical AutoCal that has already been configured is still executed, how-
ever because no user actions are necessary for this.
The measuring protection is only set if the following requirements are met:
The device is in the "Measurement" operating mode.
The inputs/outputs are not currently tested, for example, during the
commissioning/checking of communication with the control system:
→ Menu [3.11] Maintenance and diagnostics > Test
The device is not in the "Function check" operating mode.
No user is logged on.
The measuring protection is set with a delay under the following conditions:
As long as function check is active
As long as the operating mode is not "Measurement"
As long as an already started calibration is still running (including expira-
tion of the sample gas purging time)
If a calibration was triggered by a digital input, until the end of the cali-
bration or until the trigger is discarded.
If an AutoCal/AutoVal is pending and its execution is delayed
As long as an AutoCal/AutoVal is running, including sample gas inter-
mediate operation
Acknowledge current messages.
As long as the cause of a fault, a maintenance alarm, or a maintenance
demand has not been remedied, the corresponding message is still pend-
Positive edge
2)
ing.
Messages are not acknowledged in the logbook but in the message list.
Software reset
Positive edge
2)
6.8 [2.09] Settings > Inputs/outputs
.
3)
Functions
147