Siemens SIPROCESS GA700 Operating Manual page 287

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Table A- 10
Measured value status "Uncertain" (MODBUS TCP)
Quality byte
7 6 5 4 3 2 1 0
0 1 1 0 1 0 0 0 104
0 1 1 1 1 0 0 0 120
0 1 0 0 1 1 1 1 79
Table A- 11
Measured value status "Bad" (MODBUS TCP)
Quality byte
7 6 5 4 3 2 1 0
0 0 0 0 0 0 0 0 0
0 0 1 1 1 1 0 0 60
0 0 1 1 1 1 1 1 63
0 0 1 0 0 1 0 0 36
0 0 1 0 0 1 1 1 39
0 0 1 0 1 0 0 0 40
0 0 1 0 1 0 1 1 43
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Dec
Hex
Cause
0x68
The measured value is uncertain because of a
maintenance demanded.
0x78
The module is in the "Warming-up" state. The
measured value is uncertain because the ambient
conditions are outside the specifications.
0x4f
The measured value is not available because the
device is in the initialization phase (startup) or
measuring mode has not yet been started.
Dec
Hex
Cause
0x00
The measured value is invalid because it is not
available with the existing device configuration.
0x3c
The measured value is invalid for one of the follow-
ing reasons:
A calibration or an AutoCal is being carried out
Parameters are being downloaded
0x3f
The measured value is invalid for one of the follow-
ing reasons:
Firmware is being downloaded
The device is in the pause or standby state
0x24
The measured value is invalid because of a fault or
a maintenance alarm.
0x27
The measured value is invalid because of a fault or
a maintenance alarm.
0x28
The measured value is invalid because of a fault
whose cause can be found in the process (gas
supply).
0x2b
The measured value is invalid because of a fault
whose cause can be found in the process (gas
supply).
Appendix
A.5 MODBUS TCP
Effect/remedy
The measured value is still
output.
The measured value is still
output.
An initial value is output
instead of the measured
value.
Effect/remedy
The substitute value 0.0 is
output instead of the meas-
ured value.
The measured value is still
output.
A substitute value or the
last valid measured value is
output instead of the meas-
ured value.
The measured value is still
output.
A substitute value or the
last valid measured value is
output instead of the meas-
ured value.
The measured value is still
output.
A substitute value or the
last valid measured value is
output instead of the meas-
ured value.
287