Address
Name / description
0x001A
Component3_MeasValue
0x001B
0x001C
Component3_Status
0x001D
0x001E
Component3_Factor
0x001F
0x0020
Component3_Offset
0x0021
0x0022
RESERVED
0x0023
RESERVED
0x0024
Component4_Code
0x0025
Component4_UnitCode
0x0026
Component4_MeasValue
0x0027
0x0028
Component4_Status
0x0029
0x002A
Component4_Factor
0x002B
0x002C
Component4_Offset
0x002D
0x002E
RESERVED
0x002F
RESERVED
0x0030
Component5_Code
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Register type
Input / holding
Input / holding
1)
Input / holding
Input / holding
Input / holding
Input / holding
Input / holding
Input / holding
1)
Input / holding
Input / holding
Input / holding
R/W
Type
R
FLOAT32
R
UINT32
R
FLOAT32
R
FLOAT32
R
UINT8
R
UINT16
R
FLOAT32
R
UINT32
R
FLOAT32
R
FLOAT32
R
UINT8
Appendix
A.5 MODBUS TCP
Coding
DECIMAL
Bit-Enum
DECIMAL
DECIMAL
Table A-16
MODBUS TCP: Coding
of process value type
[Object ID 0x9F, 0xA1,
0xA3, 0xA5, 0xA7,
0x003C, 0x0049,
0x0054, 0x0060,
0x006C, 0x0078]
(Page 299)
Table A-17
MODBUS TCP: Coding
of process value units
[Object ID 0xA0, 0xA2,
0xA4, 0xA6, 0xA9,
0x003D, 0x004A,
0x0055, 0x0061,
0x006D, 0x0079]
(Page 299)
DECIMAL
Bit-Enum
DECIMAL
DECIMAL
Table A-16
MODBUS TCP: Coding
of process value type
[Object ID 0x9F, 0xA1,
0xA3, 0xA5, 0xA7,
0x003C, 0x0049,
0x0054, 0x0060,
0x006C, 0x0078]
(Page 299)