Siemens SIPROCESS GA700 Operating Manual page 270

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Alarm, error, and system messages
7.4 CALOMAT 7 message list
Message
No.
Text
29 to
Zero drift MR1 to MR4
32
outside tolerance
34
Analyzer module temper-
ature not in permissible
range
35
Measurement block tem-
perature uncertain
36
Membrane temperature
too high
37
Membrane temperature
too low
64
Power supply 3.3 V out-
side tolerance
65
Operating voltage 24 V
outside tolerance
66
Power supply +5 V out-
side tolerance
270
Cause
Gas line dirty, gas cooler faulty, ...
Gas lines leaking
Incorrect zero gas or setpoint faulty
Device is operated outside the permit-
ted ambient temperature range
Temperature sensor defective
Gas temperature too high
Flow rate too low
Membrane temperature control defec-
tive
Membrane defective
Strong cooling due to extremely cold
gas
Flow rate much too high compared
with the specification
Electronic fault
Electronic fault
Power supply unit fault
Electronic fault
Remedy
Calibrate zero point and full-scale
value
Check the setpoint for the zero gas
calibration in the menu "[2.08.1.1]
Setpoints"
Check the zero gas supply
Check the ambient temperature of the
analyzer
Check the ambient temperature of the
analyzer
In the menu "[3.05.1.3] Heating con-
troller", check the temperatures of the
measurement block and sensor
Check the ambient temperature of the
analyzer
Check the temperatures of mem-
branes in the menu "[3.05.1.1] Meas-
ured signal".
Check the external gas flow rate with
a suitable measuring device
Check the ambient temperature of the
analyzer
Check the temperatures of mem-
branes in the menu "[3.05.1.1] Meas-
ured signal".
Check the external gas flow rate with
a suitable measuring device
Check the plug-in connections be-
tween the power supply unit and the
module
If necessary, replace the module
The power supply unit is defective
Replace the basic device
Check the plug-in connections be-
tween the power supply unit and ana-
lyzer module
If necessary, replace the analyzer
module
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05