Executing normalization
Note
Measured value errors
If you change the setpoints but do not normalize the module, the changed values are saved
and the measured value is calculated incorrectly.
After you have changed one or both setpoint settings, always run a normalization!
1. Main menu > "2. Settings" > "20. Service > "03. Factory calibrations" > "03.
Normalization".
The following current values are displayed in the corresponding display fields:
– Factor for determining the intensity: "k norm intensity" display field
– S
– Factor for determining C
– The current phase: "Phase" display field
2. Enter setpoints:
– Authorize access with Expert PIN, if necessary.
– Enter setpoints: Use the parameter fields "Setpoint zero value" and "Setpoint end
3. Start normalization calibration wizard: "Start normalization" parameter field. After the
wizard has run, the following newly determined values are displayed:
– "k norm intensity"
– "S norm offset"
– "k norm span"
– "Phase"
6.14.2.5

[2.20.03] Factory calibrations CALOMAT 7

Executing normalization calibration
Note
Right to change setpoints
The analyzer module works with pure gases and is matched to the local application in the
factory. Make sure that the settings for the setpoints are only changed by suitably trained
personnel.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
measured value at minimum signal: "S norm offset" display field
norm
value".
: "k norm span" display field
norm
Functions
6.14 [2.20] Settings > Service