Testing Analog Inputs - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Functions
6.24 [3.11] Maintenance & Diagnostics > Test
5. Enter setpoints for additional analog outputs, if necessary: Repeat steps 2 to 4.
6. Enable/disable test mode:
Main menu > "3. Maintenance & diagnostics" > "11. Test" > "1. Test inputs/outputs" >
"1. Analog outputs" > "1. Test mode" > "Test mode" parameter box:
– When you enable test mode, the analog outputs can only be used for test purposes.
– Disable the function directly after the test is complete.
7. Measure/check test signal externally.
8. Disable test mode:
Main menu > "3. Maintenance & diagnostics" > "11. Test" > "1. Test inputs/outputs" >
"1. Analog outputs" > "1. Test mode" > "Test mode" parameter box
When you disable test mode, the analog outputs work once again according to their
parameter assignment.
See also
LUI symbol sets (Page 255)
6.24.2.2

Testing analog inputs

Requirement
Note
Availability of the menu
This menu is only available if your device has at least one of the option modules (2.1 and/ or
2.2) installed.
● Make sure that a suitable test device is connected.
● When entering the setpoint using internal analog outputs: The analog outputs of the used
option module must be connected to the analog inputs of the option module 2.
● When entering the setpoint using external analog outputs: Current signal must be applied
externally. Make sure that the analog inputs are wired accordingly.
238
This operational status can only be recognized by the function control indicators in the
status bar
(Symbol "Hand, 2 points") or
Function check indicators (Page 255).
(Symbol "Screw wrench" → Table 7-1
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05