Message
No.
Text
67
Supply voltage -5 V out-
side tolerance
69
Power supply +23 V out-
side tolerance
70
Reference voltage 2.5 V
outside tolerance
75
Temperature sensor on
AM PCB faulty
76
Temperature sensor on
measurement block de-
fective
83
Signal acquisition faulty
84
Test signal outside toler-
ance
85
System error
86
System error
87
AM parameter memory
faulty
88
Analog output faulty
89
ADC faulty
94
Membrane defective
96
Temperature sensor
PT1000 defective
128
SPI communication faulty
129
Communication with ADC
faulty
144
System error
145
Watchdog reset
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
Cause
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Electronic fault
•
Analyzer module / option module
•
faulty
Analyzer module electronics defective •
•
Internal software error
•
If the error is signaled by the AM:
•
Analyzer module electronics defective
Internal software error (analyzer
•
module or processing module)
Alarm, error, and system messages
7.4 CALOMAT 7 message list
Remedy
Check the plug-in connections be-
•
tween the power supply unit and ana-
lyzer module
If necessary, replace the analyzer
•
module
Replace the analyzer module
•
Replace the module
•
Replace the analyzer module printed
•
circuit board or the analyzer module
Replace the sensor module
•
Replace the analyzer module
•
Replace the analyzer module
•
Restart the device
•
Restart the device
•
Replace the module
•
Check the electrical circuit in the
•
menu "[3.1.11.1] Test analog out-
puts".
If necessary, replace the analyzer
•
module
Replace the module
•
Replace the sensor module
•
Replace the sensor module
•
Replace the module
•
Replace the analyzer module
Restart the device
•
Acknowledge the message.
•
Contact Service
•
271