Siemens SIPROCESS GA700 Operating Manual page 266

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Mes-
sage
No.
Text
29 to
Zero drift MR1 to MR4
32
outside tolerance.
33
Measured value greater
than full-scale value.
34
Analyzer module temper-
ature not in permissible
range.
64
Power supply 3.3 V out-
side tolerance.
65
Operating voltage 24 V
outside tolerance.
66
Power supply +5 V out-
side tolerance.
67
Supply voltage -5 V out-
side tolerance
68
Source/solenoid voltage
outside tolerance.
69
Power supply +23 V out-
side tolerance.
70
Reference voltage 2.5 V
outside tolerance.
71
Temperature sensor on
probe defective.
72
Temperature sensor on
measuring chamber
faulty.
73
Temperature sensor on
chopper faulty.
266
Cause
Gas lines contaminated
Gas cooler defective, ....
Gas lines leaking
Incorrect zero gas or setpoint faulty
Sample gas concentration too high
Sample gas pressure too high / out-
side pressure correction range
Device is operated outside the permit-
ted ambient temperature range
Electronic fault
Electronic fault
Power supply unit fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Remedy
Check the setpoint for the zero gas
calibration in the menu "[2.08.1.1]
Setpoints"
Check the zero gas supply.
U7/O7: Check the reference gas
supply.
Check the sample gas concentration
and the sample gas pressure. If nec-
essary, carry out a span gas calibra-
tion.
Check the ambient temperature of the
analyzer
Check the plug-in connections be-
tween the power supply unit and the
module
If necessary, replace the module.
The power supply unit is defective.
Replace the basic device.
Check the plug-in connections be-
tween the power supply unit and ana-
lyzer module
If necessary, replace the analyzer
module.
Check the plug-in connections be-
tween the power supply unit and ana-
lyzer module
If necessary, replace the analyzer
module.
Replace the analyzer module.
Replace the analyzer module.
Replace the module.
Replace the analyzer module.
Replace the analyzer module.
Replace the analyzer module.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05