Ultramat 7 Application Scenarios - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Table 6- 41
OXYMAT 7: Application 5
Description of application 5
Conditions
Sequence
Result
6.9.3.2

ULTRAMAT 7 application scenarios

Table 6- 42
ULTRAMAT 7: Application scenario 1
Description of application scenario 1
Conditions
Consequences
Result
Corresponds to application scenario 2 with OXYMAT 7
1)
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
The interference gas concentration is variable.
The influence of the interference gas is not known and will be determined by the wizard.
Selection of measured value source for the interference gas.
The interference gas is measured in the selected unit.
Enter substitute value in the selected interference gas unit: If the measured value of the inter-
ference gas concentration is invalid or not available, the device works with the concentration
of the substitute value.
Determine the correction coefficient k1 by means of a measuring sequence. Enter the set-
point required in the measurement procedure in the unit of the sample gas.
The influence of the interference gas is determined from the measured sample gas concentration
and the applied interference gas with the specified setpoint.
Case 1: The interference gas concentration is measured and is valid.
Case 2: The interference gas concentration cannot be measured or is invalid.
1)
The interference gas concentration is known and is assumed to be constant.
The influence of the interference gas is known
Enter the constants in the unit of the measured value.
The specified constant is subtracted from the measured value.
6.9 [2.10] Settings > Correction of cross-interferences
Functions
163

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents