Enabling/Disabling All Corrections Of Cross-Interferences; [2.11] Settings > Pressure Sensor Selection; Overview - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

4. Change the process tag assignment:
– Check the displayed process tag assignment.
– "Use for process tag" > "Change" parameter field
– Activate the corresponding process tags that are to be integrated into the correction of
5. Enabling/disabling all corrections of cross-interferences (Page 179)
6.9.9

Enabling/disabling all corrections of cross-interferences

Procedure
1. Component > Main menu > "2 Settings" > "10 Correction" > "1 Total"
2. "All corrections" parameter field:
– "On": Only the corrections enabled in the corresponding parameter assignment are
– "Off": None of the configured corrections is active.
6.10
[2.11] Settings > Pressure sensor selection
6.10.1

Overview

OXYMAT 7/ULTRAMAT 7
To correct the influence of pressure on the measured value, each OXYMAT 7 or ULTRAMAT
7 analyzer module can be equipped with an internal pressure sensor. The pressure
correction with the device-internal sensor is then preset in the factory.
Alternatively, the measured value can be corrected with external process values that are
provided by another SIPROCESS GA700 device via MODBUS TCP or via an analog input
(option module 2).
With OXYMAT 7 devices the sensor is integrated in the analyzer module. The sample gas
pressure is measured directly using the reference gas infeed. Fluctuations in the sample gas
pressure can be corrected in the following areas, depending on the installed type of pressure
sensor:
● 0.5 to 1.5 bar (absolute) → "With tubes" (SENSOR_1500_HPA)
● 0.5 to 2.5 bar (absolute) → "With pipes" (SENSOR_2500_HPA)
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
The displayed editing window includes all available process tags.
cross-interference. Deactivate process tags that are not used.
considered as part of the process tag switchover.
6.10 [2.11] Settings > Pressure sensor selection
Functions
179