Measured Value Status; Messages - Siemens SIPROCESS GA700 Operating Manual

Continuous gas analysis, operating with the local user interface
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Alarm, error, and system messages

7.6 Measured value status

7.6
Measured value status
7.6.1

Messages

Overview
The tables below include information that can be displayed for the measured value status by
the device.
Table 7- 9
Display of measured value status "Good"
Measured
Cause
value qualifier
141
The measured value is valid, but is below an alarm limit.
142
The measured value is valid, but is above an alarm limit.
143
The measured value is valid, but is not in the permitted alarm
limit range.
137
The measured value is valid, but is below a warning limit.
138
The measured value is valid, but is above a warning limit.
139
The measured value is valid, but is within or outside a warning
limit range.
188
The measured value is valid, but the function check is active.
For example, measuring mode is active and a user is logged on.
168
The measured value is valid, but a maintenance demanded is
signaled.
164
The measured value is valid, but a maintenance required is
signaled.
128
The measurement is valid (normal measuring mode).
276
Effect/remedy
-
-
-
-
-
-
-
-
-
-
Operating with the Local User Interface