4. If necessary, change assignment of process tags for correction of cross-interference
Note
Setting instruction
You can only change the assignment of process tags for correction of cross-interference
if process tag switching is enabled. To do this, proceed as follows:
• Check the displayed process tag assignment.
• "Process tag" > "Change" parameter field
• Activate the corresponding process tags that are to be integrated into the correction of
5. Enabling/disabling all corrections of cross-interferences (Page 179).
6.9.8.3

Assigning variable correction of cross-interference

Procedure with CALOMAT 7
1. Component > Main menu > "2. Settings" > "10. Correction of cross interference" > "3 ... 6
..."
2. Enable/disable correction of cross-interference: "Correction of cross-interference"
parameter field. Authorize access with appropriate Expert PIN, if necessary.
The correction of cross-interference is activated/deactivated for a specific correction
value.
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
The displayed editing window includes all available process tags.
cross-interference. Deactivate process tags that are not used.
6.9 [2.10] Settings > Correction of cross-interferences