Centrifuge Module Interface Operating Instructions
Term
Light tower
LineMaster
LIS
Load delay
Module
Offline
092D0001-02 Rev. A
Definition or Usage
Vertically stacked colored lights that indicate system
status information. The color of an illuminated light
indicates the operating status of a robot or gate.
Red indicates a serious error occurred on the
automation system.
The conveyor, robots, and instruments and
modules stop. The operator must take corrective
action.
Yellow indicates a station is offline.
The conveyor continues to run. The operator
must take corrective action.
White indicates all stations are online.
The conveyor continues to run. The operator
should take corrective action at the station.
Green indicates the conveyor is running.
Application that allows the operator to control sample
management for the entire automation system.
The LineMaster workspace is the primary point of
interaction between the operator and the automation
system.
Laboratory Information System
Time in seconds that the Centrifuge module robot
waits between the arrival of a sample tube in the
module and the placement of the cassette containing
it into the Hettich unit. This is dependent on the
availability of the Hettich unit.
Connected system, such as the Sample Manager, that
performs sample tube management functions but
does not analyze samples.
Instrument, module, or station is not operational.
Non-robotic interfaces reach the offline state when
the station no longer diverts sample tubes to the side
track.
Robotic interfaces reach the offline state after the
instrument or module checks all safety features and
the robot homes.