Centrifuge Module Interface Operating Instructions

Laser Safety

LASER WARNING
Do not look directly at the laser beam or at its reflection from a shiny
surface. This causes damage to the eyes. Only trained field service
personnel should perform procedures related to laser assemblies.
LASER WARNING
Do not perform a gate cleaning procedure without pushing down the
gate control button. Performing these procedures while the button is
lifted and the gate is started may expose you to laser light and could
cause damage to the eyes. Always push down the gate control button
before beginning a cleaning procedure.
All ADVIA LabCell and ADVIA WorkCell CDX systems are classified as CDRH
Class II and EN60825-1 Class 2 systems and comply with CDRH 1040.10
and IEC 60825-1: 1993+A1: 1997+A2: 2001.
Note These classifications are defined by the Food and Drug
Administration (FDA) Center for Devices and Radiological Health (CDRH)
regulations 21 CFR 1040.10 and 1040.11.
Particular caution should be taken when performing operating or
maintenance procedures near or around a divert gate, an interface gate, or
a Sample Manager (or Loader) with an open drawer.
Before operating the system, ensure that all safety covers are in place.
Table 3:
Characteristic
Maximum Power Output
Wavelength
Pulse duration
092D0001-02 Rev. A
Laser Specifications
Specification
1.2 mW
670 nm
111 µs
Safety Information