Removing A Sample Tube From The Robot Gripper - Siemens ADVIA LabCell Operating Instructions Manual

Centrifuge module interface
Table of Contents

Advertisement

Troubleshooting
4. If sample tubes are in the Centrifuge module and the system goes to
red Shutdown, bring the module interface Online.
A message prompts you to select a manual or automatic method to
remove the sample tubes.
a. If you select the automatic method, all tubes in the module are
b. If you select the manual method, you must manually remove the

Removing a Sample Tube from the Robot Gripper

BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to
Appendix A, Safety Information for recommended precautions when
working with biohazardous materials.
If the robot does not release a sample tube, you can use the robot gripper
release to remove the sample tube.
If you need to lift the robot arm to reach the sample tube, refer to Lifting
the Robot from the Hettich Unit , page 72 and Using the Z-up Feature to
Raise the Robot, page 74.
To remove the sample tube, perform these steps:
1. Press the E-stop button.
2. Open the robot enclosure door.
WARNING
Do not press the gripper release while holding the sides of the sample
tube. Holding the sample tube too close to the gripper can cause bodily
harm. Always hold the bottom of the sample tube.
3. Hold the sample tube below the gripper fingers.
page 78
returned to the track and system comes online.
Tubes are processed based on current spin status.
sample tubes and check them in according to your laboratory
procedures.
After removing all tubes, bring the Centrifuge module online.