Centrifuge Module Interface Operating Instructions

Location Override Tab

The Location Override tab of the Pre and Post Tests (Shift F5) screen
cancels a Centrifuge pre-test when the sample originates at a location
included in the Location list.
Note For this option, the LIS order must specify the location.
Figure 9:
To assign a sample tube from a specific location to not be centrifuged
before it proceeds to other tests, perform the following steps:
1. On the LineMaster workspace, select System > Pre and Post Tests
(Shift F5).
2. Select the Location Override tab.
3. At the Location area on the right, select the location ID that will not be
centrifuged prior to testing.
4. At the Pre and Post Test area on the left, in the Selected column,
select the appropriate pre-test check box.
5. To verify that a Location Override is set, select the Usage tab.
The information you selected is listed in the Usage tab.
092D0001-02 Rev. A
Pre and Post Tests (Shift F5)–Location Override Tab
System Operation
page 45