Centrifuge Module Interface Operating Instructions
9. Bring the Centrifuge module online.
For information about gate maintenance, see the
ADVIA Automation Solutions Operator's Guide.

Maintaining the Hettich Unit

Carefully read the Hettich unit documentation prior to performing any
maintenance procedures on the Hettich unit. Perform the maintenance
procedures at the intervals recommended by the manufacturer.
CAUTION
Do not leave broken tubes in the centrifuge. If a sample tube breaks
during centrifugation, sample tubes, cassettes, and the robot may remain
inside the Hettich centrifuge unit. If this occurs, see Managing Sample
Tube Breakage , page 80.
WARNING
Do not reach into the Hettich centrifuge unit through the hatch to
remove sample tubes, cassettes or buckets. Reaching into the Hettich
centrifuge unit through the hatch can cause bodily injury. Pull out the
Hettich centrifuge unit and open the lid to remove any components.
WARNING
Do not clean the Hettich centrifuge unit while power is on. Cleaning the
unit while power is on can cause injury. To prevent robot motion, ensure
that you press the E-stop before you clean the Hettich centrifuge unit.
CAUTION
Do not loosen or remove the rotors for cleaning. Loosening or removing
rotors can cause system damage. If it is necessary to loosen or remove
the rotors, contact your local service provider or distributor for assistance.
CAUTION
Do not use any rotor, buckets, or cassettes that show any sign of cracks,
corrosion, or mechanical failure. Using damaged components in the
Hettich centrifuge unit can cause damage to the equipment, loss of
patient samples, and contaminate the Hettich centrifuge unit. Contact
your local support provider to replace any rotor, buckets, or cassettes that
show any sign of corrosion or mechanical failure.
092D0001-02 Rev. A
Maintenance
page 55