Centrifuge Module Interface Operating Instructions
A maximum of 25 grams (g) difference in weight is allowed between
cassettes. The Centrifuge module processes sample tubes weighing
between 5 g and 25 g.
Figure 8:
1. Scale (weigh pan of the scale)
2. Cassette
3. Cassette buffer area
4. Hettich access hatch

Universal Power Supply (UPS)

The Centrifuge module contains a Universal Power Supply (UPS) that
provides temporary battery power during power loss. The UPS ensures that
in the event of a power loss, the Centrifuge has enough power to complete
the current spin cycle.
If the power is not restored within 5 seconds, the Centrifuge module
begins a shutdown process. The UPS maintains power to the Centrifuge
module to continue operations until the battery power depletes or power is
restored.
If a spin is in progress, the spin completes. After a spin completes, if the
UPS has at least 40% battery power, the robot continues operations
removing all cassettes from the Hettich unit before transitioning to
shutdown.